User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: PM3375 Sub5 DATE: 07-Apr-97 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 23 NUMBER OF LINES: 298 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 IEEE [SRQ OFF] 1.002 ASK- P F 1.003 ASK+ X 1.004 HEAD {Horizontal deflection coefficients: MTB x10} 1.005 DISP Set the TG 501 MARKER SELECTOR to 5ns. 1.006 IEEE HOR MTB,FCN ON,TIM 50E-09,MGN ON[10] 1.007 IEEE [GTL] 1.008 DISP Adjust the X POS control so that the 2nd marker pulse 1.008 DISP coincides exactly with the 2nd vertical graticule line. 1.009 MEMI Enter time between 2nd and 10th vert. grad. lines in ns: 1.010 ACC 40.00ns 0.01% 1.011 MEME 1.012 MEMC nT 4.00% 5ns’div. 2.001 DISP Set the TG 501 MARKER SELECTOR to 10ns. 2.002 IEEE HOR MTB,TIM .1E-06,MGN ON[10] 2.003 IEEE [GTL] 2.004 DISP Adjust the X POS control so that the 2nd marker pulse 2.004 DISP coincides exactly with the 2nd vertical graticule line. 2.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 2.005 DISP the 2nd and 10th marker pulses coincide exactly with 2.005 DISP the UUT 2nd and 10th vertical gradicule lines. 2.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 2.007 MATH MEM = ((MEM * 0.01) * 10) + 10 2.008 ACC 10.000ns 0.001% 2.009 MEME 2.010 MEMC uT 4.0000% 10ns’div. 3.001 DISP Set the TG 501 MARKER SELECTOR to 20ns. 3.002 IEEE HOR MTB,TIM .2E-06,MGN ON[10] 3.003 IEEE [GTL] 3.004 DISP Adjust the X POS control so that the 2nd marker pulse 3.004 DISP coincides exactly with the 2nd vertical graticule line. 3.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 3.005 DISP the 2nd and 10th marker pulses coincide exactly with 3.005 DISP the UUT 2nd and 10th vertical gradicule lines. 3.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 3.007 MATH MEM = ((MEM * 0.01) * 20) + 20 3.008 ACC 20.000ns 0.001% 3.009 MEME 3.010 MEMC uT 4.0000% 20ns’div. 4.001 DISP Set the TG 501 MARKER SELECTOR to 50ns. 4.002 IEEE HOR MTB,TIM .5E-06, MGN ON[10] 4.003 IEEE [GTL] 4.004 DISP Adjust the X POS control so that the 2nd marker pulse 4.004 DISP coincides exactly with the 2nd vertical graticule line. 4.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 4.005 DISP the 2nd and 10th marker pulses coincide exactly with 4.005 DISP the UUT 2nd and 10th vertical gradicule lines. 4.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 4.007 MATH MEM = ((MEM * 0.01) * 50) + 50 4.008 ACC 50.000ns 0.001% 4.009 MEME 4.010 MEMC nT 4.00% 50ns’div. 5.001 DISP Set the TG 501 MARKER SELECTOR to .1ęs. 5.002 IEEE HOR MTB,TIM 1E-06[10] 5.003 IEEE [GTL] 5.004 DISP Adjust the X POS control so that the 2nd marker pulse 5.004 DISP coincides exactly with the 2nd vertical graticule line. 5.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 5.005 DISP the 2nd and 10th marker pulses coincide exactly with 5.005 DISP the UUT 2nd and 10th vertical gradicule lines. 5.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 5.007 MATH MEM = ((MEM * 0.01) * .1) + .1 5.008 ACC 0.1000ęs 0.001% 5.009 MEME 5.010 MEMC nT 4.00% .1ęs’div. 6.001 DISP Set the TG 501 MARKER SELECTOR to .2ęs. 6.002 IEEE HOR MTB,TIM 2E-06[10] 6.003 IEEE [GTL] 6.004 DISP Adjust the X POS control so that the 2nd marker pulse 6.004 DISP coincides exactly with the 2nd vertical graticule line. 6.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 6.005 DISP the 2nd and 10th marker pulses coincide exactly with 6.005 DISP the UUT 2nd and 10th vertical gradicule lines. 6.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 6.007 MATH MEM = ((MEM * 0.01) * .2) + .2 6.008 ACC 0.2000ęs 0.001% 6.009 MEME 6.010 MEMC nT 4.00% .2ęs’div. 7.001 DISP Set the TG 501 MARKER SELECTOR to .5ęs. 7.002 IEEE HOR MTB,TIM 5E-06[10] 7.003 IEEE [GTL] 7.004 DISP Adjust the X POS control so that the 2nd marker pulse 7.004 DISP coincides exactly with the 2nd vertical graticule line. 7.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 7.005 DISP the 2nd and 10th marker pulses coincide exactly with 7.005 DISP the UUT 2nd and 10th vertical gradicule lines. 7.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 7.007 MATH MEM = ((MEM * 0.01) * .5) + .5 7.008 ACC 0.5000ęs 0.001% 7.009 MEME 7.010 MEMC uT 4.00% .5ęs’div. 8.001 DISP Set the TG 501 MARKER SELECTOR to 1ęs. 8.002 IEEE HOR MTB,TIM 10E-06[10] 8.003 IEEE [GTL] 8.004 DISP Adjust the X POS control so that the 2nd marker pulse 8.004 DISP coincides exactly with the 2nd vertical graticule line. 8.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 8.005 DISP the 2nd and 10th marker pulses coincide exactly with 8.005 DISP the UUT 2nd and 10th vertical gradicule lines. 8.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 8.007 MATH MEM = ((MEM * 0.01) * 1) + 1 8.008 ACC 1.0000ęs 0.001% 8.009 MEME 8.010 MEMC uT 4.00% 1ęs’div. 9.001 DISP Set the TG 501 MARKER SELECTOR to 2ęs. 9.002 IEEE HOR MTB,TIM 20E-06[10] 9.003 IEEE [GTL] 9.004 DISP Adjust the X POS control so that the 2nd marker pulse 9.004 DISP coincides exactly with the 2nd vertical graticule line. 9.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 9.005 DISP the 2nd and 10th marker pulses coincide exactly with 9.005 DISP the UUT 2nd and 10th vertical gradicule lines. 9.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 9.007 MATH MEM = ((MEM * 0.01) * 2) + 2 9.008 ACC 2.0000ęs 0.001% 9.009 MEME 9.010 MEMC uT 4.00% 2ęs’div. 10.001 DISP Set the TG 501 MARKER SELECTOR to 5ęs. 10.002 IEEE HOR MTB,TIM 50E-06[10] 10.003 IEEE [GTL] 10.004 DISP Adjust the X POS control so that the 2nd marker pulse 10.004 DISP coincides exactly with the 2nd vertical graticule line. 10.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 10.005 DISP the 2nd and 10th marker pulses coincide exactly with 10.005 DISP the UUT 2nd and 10th vertical gradicule lines. 10.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 10.007 MATH MEM = ((MEM * 0.01) * 5) + 5 10.008 ACC 5.0000ęs 0.001% 10.009 MEME 10.010 MEMC uT 4.00% 5ęs’div. 11.001 DISP Set the TG 501 MARKER SELECTOR to 10ęs. 11.002 IEEE HOR MTB,TIM .1E-03[10] 11.003 IEEE [GTL] 11.004 DISP Adjust the X POS control so that the 2nd marker pulse 11.004 DISP coincides exactly with the 2nd vertical graticule line. 11.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 11.005 DISP the 2nd and 10th marker pulses coincide exactly with 11.005 DISP the UUT 2nd and 10th vertical gradicule lines. 11.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 11.007 MATH MEM = ((MEM * 0.01) * 10) + 10 11.008 ACC 10.000ęs 0.001% 11.009 MEME 11.010 MEMC uT 4.00% 10ęs’div. 12.001 DISP Set the TG 501 MARKER SELECTOR to 20ęs. 12.002 IEEE HOR MTB,TIM .2E-03[10] 12.003 IEEE [GTL] 12.004 DISP Adjust the X POS control so that the 2nd marker pulse 12.004 DISP coincides exactly with the 2nd vertical graticule line. 12.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 12.005 DISP the 2nd and 10th marker pulses coincide exactly with 12.005 DISP the UUT 2nd and 10th vertical gradicule lines. 12.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 12.007 MATH MEM = ((MEM * 0.01) * 20) + 20 12.008 ACC 20.000ęs 0.001% 12.009 MEME 12.010 MEMC uT 4.00% 20ęs’div. 13.001 DISP Set the TG 501 MARKER SELECTOR to 50ęs. 13.002 IEEE HOR MTB,TIM .5E-03[10] 13.003 IEEE [GTL] 13.004 DISP Adjust the X POS control so that the 2nd marker pulse 13.004 DISP coincides exactly with the 2nd vertical graticule line. 13.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 13.005 DISP the 2nd and 10th marker pulses coincide exactly with 13.005 DISP the UUT 2nd and 10th vertical gradicule lines. 13.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 13.007 MATH MEM = ((MEM * 0.01) * 50) + 50 13.008 ACC 50.000ęs 0.001% 13.009 MEME 13.010 MEMC uT 4.00% 50ęs’div. 14.001 DISP Set the TG 501 MARKER SELECTOR to .1ms. 14.002 IEEE HOR MTB,TIM 1E-03[10] 14.003 IEEE [GTL] 14.004 DISP Adjust the X POS control so that the 2nd marker pulse 14.004 DISP coincides exactly with the 2nd vertical graticule line. 14.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 14.005 DISP the 2nd and 10th marker pulses coincide exactly with 14.005 DISP the UUT 2nd and 10th vertical gradicule lines. 14.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 14.007 MATH MEM = ((MEM * 0.01) * .1) + .1 14.008 ACC 0.1000ms 0.001% 14.009 MEME 14.010 MEMC uT 4.00% .1ms’div. 15.001 DISP Set the TG 501 MARKER SELECTOR to .2ms. 15.002 IEEE HOR MTB,TIM 2E-03[10] 15.003 IEEE [GTL] 15.004 DISP Adjust the X POS control so that the 2nd marker pulse 15.004 DISP coincides exactly with the 2nd vertical graticule line. 15.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 15.005 DISP the 2nd and 10th marker pulses coincide exactly with 15.005 DISP the UUT 2nd and 10th vertical gradicule lines. 15.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 15.007 MATH MEM = ((MEM * 0.01) * .2) + .2 15.008 ACC 0.2000ms 0.001% 15.009 MEME 15.010 MEMC uT 4.00% .2ms’div. 16.001 DISP Set the TG 501 MARKER SELECTOR to .5ms. 16.002 IEEE HOR MTB,TIM 5E-03[10] 16.003 IEEE [GTL] 16.004 DISP Adjust the X POS control so that the 2nd marker pulse 16.004 DISP coincides exactly with the 2nd vertical graticule line. 16.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 16.005 DISP the 2nd and 10th marker pulses coincide exactly with 16.005 DISP the UUT 2nd and 10th vertical gradicule lines. 16.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 16.007 MATH MEM = ((MEM * 0.01) * .5) + .5 16.008 ACC 0.5000ms 0.001% 16.009 MEME 16.010 MEMC mT 4.00% .5ms’div. 17.001 DISP Set the TG 501 MARKER SELECTOR to 1ms. 17.002 IEEE HOR MTB,TIM 10E-03[10] 17.003 IEEE [GTL] 17.004 DISP Adjust the X POS control so that the 2nd marker pulse 17.004 DISP coincides exactly with the 2nd vertical graticule line. 17.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 17.005 DISP the 2nd and 10th marker pulses coincide exactly with 17.005 DISP the UUT 2nd and 10th vertical gradicule lines. 17.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 17.007 MATH MEM = ((MEM * 0.01) * 1) + 1 17.008 ACC 1.0000ms 0.001% 17.009 MEME 17.010 MEMC mT 4.00% 1ms’div. 18.001 DISP Set the TG 501 MARKER SELECTOR to 2ms. 18.002 IEEE HOR MTB,TIM 20E-03[10] 18.003 IEEE [GTL] 18.004 DISP Adjust the X POS control so that the 2nd marker pulse 18.004 DISP coincides exactly with the 2nd vertical graticule line. 18.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 18.005 DISP the 2nd and 10th marker pulses coincide exactly with 18.005 DISP the UUT 2nd and 10th vertical gradicule lines. 18.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 18.007 MATH MEM = ((MEM * 0.01) * 2) + 2 18.008 ACC 2.0000ms 0.001% 18.009 MEME 18.010 MEMC mT 4.00% 2ms’div. 19.001 DISP Set the TG 501 MARKER SELECTOR to 5ms. 19.002 IEEE HOR MTB,TIM 50E-03[10] 19.003 IEEE [GTL] 19.004 DISP Adjust the X POS control so that the 2nd marker pulse 19.004 DISP coincides exactly with the 2nd vertical graticule line. 19.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 19.005 DISP the 2nd and 10th marker pulses coincide exactly with 19.005 DISP the UUT 2nd and 10th vertical gradicule lines. 19.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 19.007 MATH MEM = ((MEM * 0.01) * 5) + 5 19.008 ACC 5.0000ms 0.001% 19.009 MEME 19.010 MEMC mT 4.00% 5ms’div. 20.001 DISP Set the TG 501 MARKER SELECTOR to 10ms. 20.002 IEEE HOR MTB,TIM .1E+00[10] 20.003 IEEE [GTL] 20.004 DISP Adjust the X POS control so that the 2nd marker pulse 20.004 DISP coincides exactly with the 2nd vertical graticule line. 20.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 20.005 DISP the 2nd and 10th marker pulses coincide exactly with 20.005 DISP the UUT 2nd and 10th vertical gradicule lines. 20.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 20.007 MATH MEM = ((MEM * 0.01) * 10) + 10 20.008 ACC 10.000ms 0.001% 20.009 MEME 20.010 MEMC mT 4.00% 10ms’div. 21.001 DISP Set the TG 501 MARKER SELECTOR to 20ms. 21.002 IEEE HOR MTB,TIM .2E+00[10] 21.003 IEEE [GTL] 21.004 DISP Adjust the X POS control so that the 2nd marker pulse 21.004 DISP coincides exactly with the 2nd vertical graticule line. 21.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 21.005 DISP the 2nd and 10th marker pulses coincide exactly with 21.005 DISP the UUT 2nd and 10th vertical gradicule lines. 21.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 21.007 MATH MEM = ((MEM * 0.01) * 20) + 20 21.008 ACC 20.000ms 0.001% 21.009 MEME 21.010 MEMC mT 4.00% 20ms’div. 22.001 DISP Set the TG 501 MARKER SELECTOR to 50ms. 22.002 IEEE HOR MTB,TIM .5E+00[10] 22.003 IEEE [GTL] 22.004 DISP Adjust the X POS control so that the 2nd marker pulse 22.004 DISP coincides exactly with the 2nd vertical graticule line. 22.005 DISP Adjust the TG 501 VARIABLE TIMING control so that the 22.005 DISP the 2nd and 10th marker pulses coincide exactly with 22.005 DISP the UUT 2nd and 10th vertical gradicule lines. 22.006 MEMI Use keyboard to enter TG 501 VARIABLE TIMING reading: 22.007 MATH MEM = ((MEM * 0.01) * 50) + 50 22.008 ACC 50.000ms 0.001% 22.009 MEME 22.010 MEMC mT 4.00% 50ms’div. 23.001 END